Averaging, 2 average value / max. and min. monitoring time – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 144

Advertising
background image

IM 11M12D01-01E

8-18

8.5.2 Average Value / Max. and Min. Monitoring Time

This instrument can compute and display Average, Max. and Min. Oxygen concentra-
tions in a specified time interval. The Average, Max. and Min. can apply to each
channel and to averaging outputs a, b and c. Here we explain how to set the time
interval for averaging, and for monitoring for Max./Min.

<Setting Procedure>
1) From the panel basic display touch [Setup] key, and the [Execution/Setup] display

appears.

2) Select [Commissioning] and the Commissioning display appears.
3) Select [Others] then [Averaging]. The display of Fig. 8.19 appears.
4)Select "Set period over which average is calculated" and numerical entry display

appears. To set 3 hours, enter 003. Entry range is 1 to 255 hours.

5)Select "Set period over which maximum and minimum is stored" and numerical entry

display appears. To set 48 hours, enter 048. Entry range is 1 to 255 hours.

<Default setting>
[Period over which average is calculated] is set to 1hr, and [Period over which maxi-
mum and minimum] is stored is set to 24 hrs prior factory shipment or after data
initialization.. .

Enter

F48E.EPS

Set period over which
average is calculated:
1 h

Set period over which

maximum and minimum is
stored: 2 4 h

Averaging

Fig. 8.19 Average / Max. / Min. Setup