IM 11M12D01-01E

10-19

10. Other Functions

10.5 Blowback

This section explains the parameter settings for performing blowback.

10.5.1 Mode

There are three modes of blowback operation: no function, semi-automatic, and auto-
matic. Blowback is not performed when the mode is set to No Function. In Semi-auto
mode, blowback can be started by key operation on the display or by a contact input
signal, and then sequentially performed at a preset blowback time and hold time. In
Auto mode, blowback is automatically performed at preset intervals. For Semi-auto or
Auto modes, blowback is performed one channel at a time until all channels have been
covered. The following restrictions apply:

* When [No function] is selected:

Blowback is not performed

* When [Semi-auto] is selected:

Semi-auto blowback can be performed. (Blowback does not start at Auto blowback start
time.)

* When [Auto] is selected:

Blowback can be performed in either Auto or Semi-auto mode.

<Setting Procedure>
1) From the basic panel display, touch Setup key; on the Execution/Setup display which

appears, select [Maintenance].

2) On the [Maintenance] display, select [Blow back setup] and the Mode selection pull-

down allows you to select between [No function], [Semi-auto] and [Auto] (see Fig.
10.23).

Enter

F75E.EPS

Mode: Auto

Hold time: 1 0 min 0 0 s

Blow back time:

3 0 min 0 0 s

Blow back setup

No function

Semi-auto
Auto

Fig. 10.23 Mode Selection