IM 11M12D01-01E

10-12

10.4.2 Procedure for Performing an Indication Check

You can perform two-point Span and Zero (Gas) Indication Checks, you can include a
third gas and perform a three-point indication check, or you can check just one of the
three points.

<Setting Procedure>
Select [Points] on the Indication check setup display. A dropdown menu opens, allowing
you to select the desired indication check point(s) from: [Span-Zero], [Span], [Zero],
[Span-Zero-Third] or [Third] (Figure 10.10).

Enter

F63E.EPS

Points: Semi-auto-Third

Timing

Indication check setup

Mode: Semi-auto

Span-Zero

Span
Zero
Span-Zero-Third
Third

Figure 10.10 Indication Check Point Selection

10.4.3 Setting Contacts for Operating Third Check Gas Solenoid Valve

When performing a three-point Semi-Auto or Auto Indication Check, you need to
allocate a contact to operate the third check gas solenoid valve.

<Setting Procedure>
1) On the basic panel display, if you touch the Setup key, the [Execution/Setup] display

appears. Select [Commissioning] then [Contact Setup].

2) Select a contact that is not yet allocated to other functions, and select [Others].
3) Select [Third gas] and turn it [ON] .

Enter

Warm up:

OFF

Range change:

OFF

Maintenance:

OFF

Calibration: OFF

Indication check: OFF

Blow back:

OFF

Third gas:

OFF

F10.11E.EPS

ON

Contact3

Others

OFF

Figure 10.11 Allocating Third Check Gas Contact