5.2.3

Power and Ground Wiring

Connect the power wiring to the L and N terminals in the averaging converter. Use
wires with a size of 2 mm2 (14 AWG) or larger. If a cable gland is not installed on the
wiring hole of the averaging converter, use a wire with an outside diameter of 17 mm or
smaller. If installed, use a wire with an outside diameter of 6 to 12 mm. A lock washer
must be installed.

Connect the ground wiring to the protective ground terminal in the averaging converter
or the external ground terminal on the side of the case. Run the ground wiring so the
ground resistance is 100

or less (equivalent to Class D grounding).

Make sure that the jumper plate is connected between the G terminal and the protective
ground terminal.

Grounding to the ground terminal

on the Averaging converter case

Converter Case

FG Terminal

Lock Washer

Crimp Terminal

Figure 5.6 Grounding to the Ground Terminal

WARNING

Operating the instrument with the jumper plate disconnected may result in poor

resistance to noise.

5.2.4

Power Wiring to Detector Heaters

This wiring provides electric power from the averaging converter to the heater for
heating the sensor in a detector. Use 2-core shielded PVC insulated PVC sheathed
control cables. The wire size should be 1.25 mm2 (16 AWG) or larger and the conductor
two-way resistance should be 10

or less. If a cable gland is not installed on the wiring

hole of the averaging converter, use a wire with an outside diameter of 17 mm or
smaller. If installed, use a wire with an outside diameter of 6 to 12 mm. The shields
should be connected to cable shield ground terminals on the case of the averaging
converter.