2 calibration setup, Caution, Calibration setup – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 155: 1 mode

Advertising
background image

IM 11M12D01-01E

9-5

9. Calibration

9.2 Calibration Setup

9.2.1 Mode

There are three calibration modes (for zero and span calibration): manual mode, semi-
auto mode (started from the display or started by contact input) using preset calibration
time and stabilization (settling) time, and autocalibration at preset intervals. Restrictions
related to each mode are described here.

- When [Manual Mode] is selected

In this mode, only Manual calibration can be performed. (Contact input used to start
semi-auto calibration has no effect). Autocalibration does not start at autocalibration
start time.

- When [Semi-Auto Mode] is selected.

In this mode, manual calibration or semi-auto calibration can be performed.
Autocalibration does not start at autocalibration start time.

- When [Auto Mode] is selected.

Any calibration mode is valid.

CAUTION

Indication Check mode of [Auto] cannot coexist with [Auto] Calibration mode. There-
fore, If the mode of Indication Check is set to [Auto], you cannot set the mode of
Calibration to [Auto], and vice versa.

< Setting Procedure>
1) From the basic panel display, if you touch the Setup key, the [Execution/Setup]

display appears. Select [Maintenance].

2) From the [Maintenance] display, select [Calibration setup] then [Mode] and you can

select Manual, Semi-Auto, or Auto in the selection pulldown that appears. ( Fig. 9.4
).

Enter

F52E.EPS

Mode:

Points:

Zero gas conc: 0 . 9 8 %

Span gas conc: 2 1 . 0 0 %

Timing

Cal. history initialize

Manual

Semi-Auto
Auto

Calibration setup

Fig 9.4 Calibration mode setting.