4 stop valve (part no. l9852cb or g7016xh), 3 probe insertion hole, 6 piping for indication check – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 73

Advertising
background image

IM 11M12D01-01E

4-7

4. Piping

4.1.6

Piping for Indication check

To perform an indication check using the third check gas, the piping to allow the third
gas to flow into the calibration gas line is required for the ZA8F Flow Setting Unit in
addition to the system configuration shown in Figure 4.1.

Connect the piping from the third check gas cylinder through a pressure reducing valve
and a stop valve, along with the span gas line, to the zero gas inlet of the ZA8F Flow
Setting Unit. To perform an indication check, open the stop valve on the line from the
third gas cylinder, while open the stop valve on the span gas line to perform span point
calibration.

Span gas cylinder
(Instrument air)

Zero gas cylinder

Air Set

Pressure regulator

Instrument air

Stop valve

Flowmeter

Needle valve

Reference gas

line

Calibration gas

line

ZA8F flow setting unit

Third check gas cylinder

Reference gas

line

Calibration gas

line

ZA8F flow setting unit

Flowmeter

Needle valve

F3.eps

Figure 4.8 Piping to ZA8F Flow Setting Unit for Indication Check Using 3rd

Check Gas