IM 11M12D01-01E

5-9

5. Wiring

51 52 53 54 55 56 57 58

7 8

CH1 CH2 CH3 CH4

Shield

Averaging Converter

ZR22G / ZO21D

Detector #2

Detector #3

Detector #4

Detector #1

Basic Power Unit

Figure 5.7 Power Wiring to Detector Heaters

If the ambient temperature at the detector installation site exceeds 80˚C, install a
terminal box and connect to the detector using 600 V silicon rubber insulated glass
braided wires. The distance between the detector and the terminal box should be
minimized.

51 52 53 54 55 56 57 58

7

W

CH1 CH2 CH3 CH4

Shield

Averaging Converter

ZR22G / ZO21D

To Detector #2

To Detector #3

To Detector #4

Detector #1

Silicon Rubber Insulated
Glass Braided Wire

Terminal Box

Basic Power Unit

Figure 5.8 Power Wiring to Detector Heaters (Using Terminal Box)