7-1

7. Startup

7.

Startup

This section explains the startup procedures for the Averaging Converter:
supplying power, setting parameters as required, and performing manual
calibration.

7.1 Startup procedure

The startup procedure is as follows:

CAUTION

If you connect Model ZO21D

*

detectors, then you need to change the detector

parameters. Before connecting power, refer to: Sect. 7.4 Setting Detector Model.

Check valve type setting

Check piping & wiring

Check settings

F7.1E.eps

If connecting ZO21D

Briefly power on

Set detector to ZO21D (front panel)

Power OFF

If all detectors are ZR22G

Permanent power wiring

Start set parameters

Warmup, then calibrate