10.5.4 Operation of Blowback

Figure 10.26 shows blowback timing. To execute blowback by a contact input, use it
with an on-time period of 1 to 11 seconds. Once blowback starts, the SV-CH contact
output repeatedly opens and closes at an interval of approximately 10 seconds during the
preset blowback time. During the blowback time and hold time on a channel, its indi-
vidual analog output remains held at a value preset in the Output Hold Setup and
excluded from averaging group calculations. After the hold time has elapsed, blowback
is started on the next channel.

Ch1 output hold

Ch2 output hold

Ch1 blowback time

Ch2 blowback time

10 s

10 s

F10.26E.eps

Average output

Blowback start in

Blowback time

Ch1 output hold

Ch1 removed from Average

On

Off

Solenoid valve switching

Blowback start

Off

On

Hold time

Blowback time

Ch2 output hold

Ch2 removed from Average

Hold time

SV-Ch1 contact

SV-Ch2 contact

Fig. 10.26 Blowback Timing

<Default setting>
Blowback parameters are set as shown in Table 10.6 prior factory shipment or after data
initialization.

Table 10.6 Blowback Default Setting

T10.6E.eps

Item

Init. Setting

Mode

No function(disabled)

Hold time

3 minutes 00 second

Blowback time

3 minutes 00 second

Interval

30 day 00 hour

Start date

01/ 01/ 01

Start time

00: 00