5 blowback, 1 mode, Err2 – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 213: 3 error displays

Advertising
background image

IM 11M12D01-01E

12-3

12. Troubleshooting

12.1.3

Error Displays

By touching the error icon when an error occurs, detailed error descriptions can be
displayed. Use the Next key to move to the next page if the error descriptions are
displayed in multiple pages. On the detailed error description display, touch the Update
key to update the error occurrence condition. When an error and an alarm occur at a
time, an error icon takes precedence.

Tag:

Ch1

Err2

Ch2
0.96

%

Ch3
0.96

%

Ch4
0.96

%

Ch5
0.96

%

Ch6

Err1

Ch7
0.96

%

Ch8
0.94

%

F02-cE.EPS

Hold

Channels Where Errors Occur (displayed in red)

Error Icon

Figure 12.1 Error Displays on Multi-channel Display

Tag:
Ch1

Err2

Ave-a 0.96 %O2
Ave-b 0.95 %O2
Ave-c 0.96 %O2

Hold

F02-dE.EPS

Figure 12.2 Error Displays on Single-channel Display

Tag:

Err4 : Ctrl. card
Ch1 Err1 : Cell voltage
Ch2 Err2 : Heater temp.
Ch3 Err2 : Heater temp.
Ch4 Err1 : Cell voltage
Ch5 Err1 : Cell voltage
Ch6 Err2 : Heater temp.
Ch7 Err2 : Heater temp.

Fm-8E.EPS

Next

Tag:

Ch8 Err3 : Ch. card

Fm-9E.EPS

Previous

Figure 12.3 Error Descriptions (page 1) Figure 12.4 Error Descriptions (page 2)