5. Wiring

5.1.4

Types of Wiring and Cables

Types of wiring and cables to be used are listed in Table 5.1. The purpose of a shield is
not only to prevent the effect from noise disturbance but also to restrict the noise emitted
from the converter. Specified cables must be used.

Table 5.1 Types of Wiring and Cables

Terminal

Indication

Shield
Requirement

Number of
Cores

Requirements

Power,

internal grounding

External grounding

Detector heater

Detector signal

Individual analog

output

Averaging analog

output

Solenoid output

Individual error

contact output

Common contact

output

Common error

contact output

Contact input

L, N,

HTR1 to HTR8

CELL, TC, CJ

AO

Ave-a, Ave-b,

Av3-c

SV-COM, SV-CH1

to SV-CH8

DO-CH1 to

DO-CH8

DO1 to DO4

DO5

DI1, DI2, DI-COM

Unshielded

Unshielded

Shielded

Shielded

Shielded

Shielded

Shielded

Unshielded

Unshielded

Unshielded

Unshielded

2 or 3

2

6

2

Note

Note

Note

Note

Note

2 or 3

Size: 2 mm

2

(14 AWG) or larger per core

Either internal grounding or external

grounding will do.

Size: 1.25 mm

2

(16 AWG) or larger,

Conductor two-way resistance: †10 ‰

Conductor two-way resistance: †10 ‰

†550 ‰, including load resistance

Loop resistance when contact is closed:

†200 ‰

Note: The number of cores is determined by number of contacts to be used.