7.9.2 Performing Manual Calibration

1) Confirm that the zero gas flow setting valve is fully closed, then adjust supply gas

pressure to measurement gas pressure + 50kPa (but a maximum of 300 kPa) using
the valve on the gas cylinder.

2) Touch [Setup] on the touchdisplay to switch to the [Execution/Setup] display.
3) Touch [Calibration] and the Calibration display appears.
4) Touch [Manual calibration] then the [Enter] key. The Channel Select display shown

in Fig. 7.10 appears.

Enter

Ch1

Ch2

Ch3

Ch4

Ch5

Ch6

Ch7

Ch8

F7.10E.EPS

Select calibration channel

Manual cal.

Fig. 7.10 Channel Select for Manual Calibration

5) Select the channels to be calibrated (channels in warmup or error status can't be

selected). Here we describe the situation for Channel 1.

6) Select [Span calibration] and touch the [Enter] key. A display like Fig. 7.11 appears.

Enter

Manual cal.

Check the span gas

concentration value.

Next

F7.11E.EPS

Ch1

Cancel calibration

Change value: 2 1 . 0 0%

Fig. 7.11 Span Calibration setup

7) Touch [Next] and touch the [Enter] key. The display in Fig. 7.12 appears. If wiring

and piping for autocalibration has been done, span gas will start flowing at this point.

Enter

F7.12E.EPS

Ch1

Manual cal.

Open span gas valve.
Set flow span gas to
600ml/min.

Valve opened

Cancel calibration

Fig.7.12 Span Calibration start