3 other display-related functions, Display setup – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 171

Advertising
background image

IM 11M12D01-01E

10-9

10. Other Functions

10.3 Other Display-related Functions

10.3.1

Auto-return Time

On the Execution/Setup display or lower level menu displays, if no keys are touched for
a preset time, the "auto-return" time, then the display will automatically revert to the
basic panel display. The "auto-return" time can be set in the range 0 to 255 minutes. If it
is set to 0, then the display does not automatically revert.

<Setting procedure>
1) On the basic panel display, touch the Setup key to display the Execution/Setup

display, then select [Maintenance]. Select [Display setup] then [Auto return time].

2) The numerical entry display appears for you to enter the desired "auto-return" time.

To set an "auto-return" time of one hour, enter 060.

3) If you set 0, then the "auto-return" function does not operate.

<Default setting>
The Auto return time is set to 0 prior factory shipment or after data initialization.

10.3.2

Selecting Language

You can select a display language from among English, Japanese, German and French.
The display language is set to the one specified in the purchase order when the analyzer
is shipped from the factory.

<Setting procedure>
1) On the basic panel display, touch the Setup key to display the Execution/Setup

display.

2) Select [Maintenance], then [Display setup].
3) Select [Language] and the dropdown selection (Figure 10.7) allows you to select the

desired language.

Enter

F60E.EPS

Display setup

Trend graph

Auto return time: 0 min

Language:

Lcd auto 0 min

Display contrast

Tag name:

English

Japanese
Deutsch
Français

Figure 10.7 Selecting the Display Language