4 setting indication check timing, Danger, Warning – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 207: 2 cleaning

Advertising
background image

IM 11M12D01-01E

11-13

11.

Inspection and Maintenance

Cap

Flat-blade
screwdriver

Fuse

Socket

F11.6E.EPS

Fig 11.9

WARNING

If a replaced fuse blows out immediately, there may be a problem in the internal circuit of the

averaging converter, the detector, or the wiring between the averaging converter and the detector.

Carefully investigate the cause of fuse failure. If the cause is unknown, contact our service office.

DANGER

To work safely, turn off the power to the averaging converter at the external breaker

and position the power switch to "O (OFF)" before replacement.

After replacement, position the power switch to "I (ON)", attach the front cover, and

then apply the power to the averaging converter.

11.4.2

Cleaning

Use a dry soft clean cloth for cleaning during inspection or maintenance.