Solenoid valve contacts, Input contacts, 3 checking contact inputs – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 125

Advertising
background image

IM 11M12D01-01E

7-15

7. Startup

7.11.2 Checking contacts used to operate solenoid valves during autocalibration

1) From the Contact check display of Fig. 7.18 select [Solenoid valve contacts]. From

the corresponding Solenoid valve contact select display of Fig. 7.20 select the
solenoid to check. To check the Autocalibration gas switching valve, select [Com].
Touching the Enter key opens the [Span] (Span valve open) / [Zero] (Zero valve
open) selection window. Selecting an output and touching [Enter] allows you to test
Open / Close the corresponding solenoid valve.

2) To check a channel switching valve, select its channel [Ch1] through [Ch8] and touch

[Enter] to open the [OFF] (valve closed) / [ON] (valve open) selection window.
Selecting an output and touching [Enter] allows you to test Open / Close the corre-
sponding solenoid valve.

Note

You cannot check a Channel Select valve if the corresponding Channel unit is not

installed.

Enter

Ch1

F

Zero

Com

F X p

Ch2 : OFF

Ch3 : OFF

Ch4 : OFF

Ch5 : OFF

Ch6 : OFF

Ch7 : OFF

Ch8 : OFF

F30E.EPS

Solenoid valve contacts

Span

Fig. 7.20 Solenoid valve contact select

7.11.3 Checking Contact Inputs

1) From the Contact check display of Fig. 7.18 select [Input contacts] then touch the

Enter key. The display of Fig. 7.21 appears.

2) The input contact status [Open] or [Closed] is displayed in real time.

Enter

F31E.EPS

Input contact 1 : Open

Input contact 2 : Open

Input contacts

Fig. 7.21 Contact input check