1 contact output operation check, Contact1 – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 140

Advertising
background image

IM 11M12D01-01E

8-14

< Setting Procedure >
1) Touch the Setup key, and the [Execution/Setup] display appears.

2) From [Commissioning] select [Contact setup] and touch the Enter key.

3) Select contact no. to set up and touch the Enter key.

4) Select [Alarms] or [Others] and touch Enter key. You can set the condition(s)
allocated to contact output. (Fig. 8.14).

5) Select ON/OFF pulldown to set items ON and allocate the corresponding condition to
contact output. You can allocate multiple conditions to a contact output, then if any of
the conditions is true there will be contact output.

Enter

Low Low alarm:

Low alarm: ON

High alarm: OFF

High High alarm: OFF

Cal alarm: OFF

Cal gas low: OFF

Proc. gas alarm: OFF

Error: OFF

Internal temp: OFF

F8.14E.EPS

Contact1

Alarms

OFF

Fig. 8.14 Alarm Function Allocation

<Default setting>
Contact output setting are set to defaults in Table 8.5 prior factory shipment or after data
initialization..

Table 8.5 Contact function initial defaults

ON

ON

ON

ON

T8.5E.eps

(Under) warmup

Output range switch

(Under) calibration

(Under) indication check

(Under) maintenance

Blowback

3rd check gas

Normally Closed/Open

N.O.

N.O. N.O.

N.O.

Item

Alarms

High High Alarm

High Alarm

Low Alarm

Low Low Alarm

CAL coeff. Alarm

CAL gas pressure low

Process gas alarm

Temperature alarm

Error

Other status

Contact
output 1

Contact
output 2

Contact
output 3

Contact
output 4