11.4 Inspection and Maintenance of the Averaging Con-

verter

The averaging converter does not require routine inspection and maintenance. This
section describes inspections performed at the time of periodic maintenance service, part
replacement, and addition of channel cards.

11.4.1

Fuse Replacement

The averaging converter has fuses for detector heaters of channels CH1 to CH4 in the
basic power supply unit and fuses for detector heaters of channels CH5 to CH8 in the
expansion power supply unit. Fuses are inserted in both poles of the heater of each
channel. When the fuse blows out, follow the procedure below to replace it.

DO-CH1

NC

COM

32

31

DO1

NC

COM

12

11

49

48

47

46

45

44

43

29

28

27

26

SV-COM

SV-CH4

SV-CH3

SV-CH2

SV-CH1

DO-CH4

NO

NC

COM

42

41

40

DO-CH3

NO

NC

COM

39

38

37

DO-CH2

NO

NC

COM

36

35

34

NO

33

DO5

NO

NC

COM

25

24

23

DO4

NO

NC

COM

22

21

20

DO3

NO

NC

COM

19

18

17

DO2

NO

NC

COM

16

15

14

NO

13

58

57

56

55

54

53

52

51

HTR4

HTR3

HTR2

HTR1

FUSE

HTR2

HTR2

FUSE

FUSE

HTR1

HTR1

FUSE

FUSE

HTR4

HTR4

FUSE

FUSE

HTR3

HTR3

FUSE

250V T2.5A X 8

250V T2.5A X 8

FUSE

HTR7

HTR7

FUSE

FUSE

HTR8

HTR8

FUSE

FUSE

HTR5

HTR5

FUSE

FUSE

HTR6

HTR6

FUSE

HTR5

HTR6

HTR7

HTR8

101

102

103

104

105

106

107

108

81

82

83

COM

NC

NO

DO-CH5

84

85

86

COM

NC

NO

DO-CH6

87

88

89

DO-CH7

90

91

92

DO-CH8

SV-CH5

SV-CH6

61

62

63

64

SV-CH7

SV-CH8

68

67

66

65

69

70

71

72

NO

NC

COM

NO

NC

COM

Fuse Holder
(8 holders in basic power supply unit)

F11.8Eeps

Fuse Holder
(8 holders in expansion power supply unit)

Fig11.8

<How to replace the fuse>

1) Before replacement, turn off the power to the averaging converter at the external

breaker.

2) Remove the front cover, and position the power switch to "O (OFF)."
3) Remove the fuse from its holder. Using an appropriate size flat blade screwdriver,

turn the fuse holder cap counterclockwise until it stops. Pull out the cap. The fuse
will come out together with the cap.

4) Use a recommended fuse, shown below, or equivalent for replacement. Attach a new

fuse to the cap, place in the fuse holder, and push and turn the cap clockwise until it
stops with the flat blade screwdriver.
Fuses used in the averaging converter are as follows.
Maximum rated voltage: 250 V
Maximum rated current: 2.5 A
Type:

Time lag fuse

Compliance:UL, CSA, VDE, Japan's Electrical Appliance and Material Safety Law
Part number:A1112EF

5) It is recommended that the fuse be replaced every two years even if it has not blown.
6) Position the power switch to "I (ON)", and attach the front cover.

Advertising