3 lcd auto off, 4 display contrast adjustment, 5 tag name entry – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 204: Caution, 5 replacement of filter assembly, 6 replacement of o-ring, 7 cleaning the high-temperature probe adapter

Advertising
background image

IM 11M12D01-01E

11-10

11.3.5

Replacement of filter assembly

Set the filter assembly (1) in place using a special pin spanner (with a pin 4.5 mm in
diameter: part no. K9471UX or equivalent). If a filter assembly that has already been
replaced once is used again, apply grease (NEVER-SEEZ: G7067ZA) to the threads of

the filter assembly.

11.3.6

Replacement of O-ring

The detector uses three different types of O-rings (14), (21), and (22). One O-ring alone
(14), or two O-rings (21) and (22) are used. (For a pressure-compensating model, two O-
rings are used for individual uses. Two O-rings (21) and (22) are used for reference-gas
sealing and require periodic replacement.

T11.1E.ESP

Part No.

Description

(7) K9470BJ

Metal ring

(14) K9470ZS

Metal ring with grease

(21)

Two pairs of O-rings

(22)

with grease

K9470ZP

11.3.7

Cleaning the High-temperature Probe Adapter

CAUTION

• Do NOT subject the probe of the High-temperature Probe Adapter (ZO21P-H-A) to

shock. This probe uses silicon carbide (SiC) which may become damaged if it is

subjected to a strong shock or thermal shock.

The high-temperature detector is structured so that the gas to be measured is directed
toward the detector with the high-temperature probe adapter. Therefore, if the probe or
the sample gas outlet clogs, a precise measurement is no longer possible because of no
gas flow. If you use the high-temperature detector, you have to inspect it periodically
and, if any part of it is significantly clogged with dust, clean it.

Dust found sticking to the probe should be blown off. If any dust still remains after the
blowing, clean it with a metal rod, etc., inserted. In addition, if dust is found on the
auxiliary ejector or needle valve (choke) at the sample gas outlet, remove these parts
from the high-temperature probe adapter and then clean them. To remove dust, blow air
on them or rinse them with water.