<Setting Procedure>
This setting example shows how to set "When contact input 1 Opens, Start Semi-Auto
Calibration".
1) From the panel basic display touch [Setup] key, and the [Execution/Setup] display
appears.
2) Select [Commissioning] and the Commissioning display appears.
3) Select [Contact Setup] then [Input Contacts].
4) Select [Input 1] . Function Selection window (Fig 8.16) appears.
5) Select [Calibration Start].
6) For [Input contact 1] select (operate if) [Closed]. (You can select Open or Closed).
7) Change [Closed] to [Open].
<Default setting>
All contact inputs are set to [No function](disabled) prior factory shipment or after data
initialization..
Enter
F8.16E.EPS
Input 1 : Open
᭜
Input 1
᭜
Input 2
᭜
Input 2
Close
Input contacts
No function
Range change
Calibration start
lnd. check start
Blow back start
Cal. gas press. low
Process gas alarm
No function
Fig. 8.16 Contact Input Setting display