ZO21R-L Probe Protector

This probe protector is required for the general-use detector when it is used for oxygen
concentration measurements in powdered coal boilers or in fluidized furnaces to prevent
abrasion due to dust particles when gas flow exceeds 10 m/s.

When using a ZR22G general-use separate-type detector in the horizontal position, be
sure to select a probe protector ZO21R-L-

ᮀᮀᮀ-ᮀ*B to support the probe.

Insertion Length: 1.05 m, 1.55 m, 2.05 m.

Flange: JIS 5K 65A FF SUS304. ANSI Class 150 4 FF (without serration). However,

flange thickness is different.

Material: SUS316 (JIS), SUS304 (JIS) (Flange)

Weight: 1.05 m; Approx. 6/10 kg (JIS/ANSI), 1.55 m; Approx. 9/13 kg (JIS/ANSI),

2.05 m; Approx. 12/16 kg (JIS/ANSI)

Installation: Bolts, nuts, and washers are provided for detector, probe adapter and

process-side flange.

T2.2E.EPS

ZO21R

JIS 5K 65 FF SUS304
ANSI Class 150 4 FF SUS304

-L

Probe Protector (0 to 700

ЊC)

-100
-150
-200

1.05 m (3.5 ft)
1.55 m (5.1 ft)
2.05 m (6.8 ft)

Insertion
length

Flange ( *1)

*1 Thickness of flange depends on dimensions of flange.

Model

Suffix code

Option

code

Description

-J

-A

Style code

*B

Style B

Flange<1>

JIS 5K 65 FF
SUS304

155

228.6

130

190.5

5

12

A

B

C

t

ANSI Class
150 4 FF SUS304

40

50

D

F2.3E.EPS

D

Flange <1>
(with bolts, nuts,
gasket and washer)

gasket (t1.5)

SUS316

Dimensions of
holes on opposing surface

л〈

Gas flow

Washer (12)

Mounting nut (M12)

l (Insertion length)

l=1050,1550,2050

л

B

C

л

B

4 -

л15

8 -

л19

Unit

Ϻ mm

t

л

60.5