10. Other Functions

10.7 Parameter Initialization

Parameter settings can be initialized to the factory default settings. Initialization can be
done for all parameters or for individual parameters. The parameters that can be initial-
ized and their defaults are listed in Table 10.7.

<Initialization procedure>
1) On the basic panel display, touch the Setup key to display the Execution/Setup

display.

2) Select [Commissioning], next [Others] then [Defaults]. A display like Figure 10.32.1

appears.

3) Select the desired item to be initialized then a display like Figure 10.32.3 appears.

Select [Defaults start] then initialization starts.

4) When initialization completes then you are returned to a display like Figure 10.32.2.

Display

Calibration

Blow back data

mA-outputs

Alarms

Contacts data

Others

All data

Enter

Defaults

F10.32.1E.EPS

Fig. 10.32.1 Defaults Display

Enter

F10.32.2E.EPS

Defaults start

Defaults

All data

Cancel

Fig. 10.32.2 "Initialize All Data" Start Display

WARNING

Do not turn off power during initialization. Otherwise, initialization will not be per-
formed properly