IM 11M12D01-01E

5-11

5. Wiring

If the ambient temperature at the detector installation site exceeds 80˚C, install a
terminal box and connect to the detector using 600 V silicon rubber insulated glass
braided wires. The distance between the detector and the terminal box should be
minimized.

To Detector #2

ZR22G / ZO21D

CH1

CH2

Detector #1

CH3

CH4

Shield

Averaging Converter

Terminal Box

Silicon Rubber Insulated Glass Braided Wire

Channel Card

CELL

TC

CJ

ϩ
Ϫ
ϩ
Ϫ
ϩ
Ϫ

1

2

3

4

5

6

1

2

3

4

5

6

1

2

3

4

5

6

1

2

3

4

5

6

1

2

3

4

5

6

Figure 5.10 Signal Wiring to Detectors (Using Terminal Box)

5.2.6

Ground Wiring of Detectors

Connect the ground wiring to the external ground terminal on the detector case or the
internal ground terminal. Follow the instructions below. The ground terminal screw is
M4.

(1) The ground resistance should be 100

or less (equivalent to Class D grounding).

(2) If the ambient temperature of the wiring installation exceeds 80˚C, use appropriate

heat resistant wires.

WARNING

The flange bolts alone do not provide durable grounding of the detector. Be sure to

grounded the ground terminal.