7-8

7.6

Confirmation of Detector Type Setting

Check that the detector in Figure 7.7 is the one for this equipment.

WARNING

• If this converter is to be used in conjunction with the ZO21D*, the power require-

ments are limited to 115 V AC or less, 50 Hz or 60 Hz (it cannot be used with a 125
V or greater, or in the EEC).

• If detector settings are to be changed, first disconnect the wiring connections between

the detector and the converter. Then change detector settings appropriately.

* ZO21D is not a explosionproof equipment.

7.7

Selection of Measurement Gas

Combustion gases contain moisture created by burning hydrogen in the fuel. If this
moisture is removed, the oxygen concentration might be higher than before. You can
select whether the oxygen concentration in a wet gas is to be measured directly, or
compensated for its dry-gas value before use. Select the “Select measure gas: Wet” in
Figure 7.8 to select either wet or dry gas.

7.8

Output Range Setting

This section sets forth analog output range settings. For details, consult Section 8.1,
”Current Output Setting” later in this manual.

Minimum Current (4 mA) and Maximum Current (20 mA) Settings

To set the minimum and maximum current settings, follow these steps:

(1) Select the Setup from the Execution/Setup display.
(2) From the Commissioning (Setup) display, select “mA-output setup”; the display

shown in Figure 7.9 then appears.

(3) Select “mA-output1” from the “mA-outputs” display. The “mA-output1 range”

display shown in Figure 7.10 then appears.

(4) In the display shown in Figure 7.10, select “Min. oxygen conc.” and press the [Enter]

key to display the numeric-value entry display. Enter the oxygen concentration at a
4-mA output; enter [010] for a ten-percent concentration measurement.

(5) Also in Figure 7.10, select “Max. oxygen conc.” at a 20-mA output. Enter the

appropriate maximum oxygen concentration (at the 20-mA output) in the same
manner as in step 4 above.

(6) Set “mA-output2” in the same manner as in the appropriate steps above.

Both outputs are 4-20mA

mA-output1

mA-output2

Set presets

Enter

mA-outputs

F7.9E.EPS

Parameter: Oxygen
Set range

Min. oxygen conc:

0 %O2

Max. oxygen conc:

25 %O2

Output damping: 0 s

Output mode: Linear

Enter

mA-output1 range

F7.10E.EPS

Figure 7.9 Setting “mA-output” Figure 7.10 Setting “mA-output1 range”