Manual calibration – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 115

Advertising
background image

IM 11M13A01-02E

7-16

(3) Follow the display message in Figure 7.19 to turn on span gas flow. Open the span-

gas flow valve for the Flow Setting Unit by loosening the valve lock-nut and slowly
turning the valve shaft counterclockwise to flow the span gas at 600 ml/min.

±

60

ml/min. Use the calibration gas flowmeter to check the flow.

Enter

Open span gas valve.

Set flow span gas to

600ml/min.

Valve opened

Cancel calibration

Manual calibration

F7.19E.EPS

Enter

21.00%

0.5min./div

1.00%

21.0 %O2

CAL. TIME

Manual calibration

F7.20E.EPS

Figure 7.19 Span-gas Flow Display Figure 7.20 Manual Calibration Trend Graph

(for Manual Calibration)

(4) If “Valve opened” is selected as in Figure 7.19 , an oxygen-concentration trend graph

(with the oxygen concentration being measured) appears (see Figure 7.20). The CAL
TIME in the bottom area of the panel flashes. Observe the trend graph and wait until
the measured value stabilizes in the vicinity of 21 percent on the graph. At this point,
calibration has not yet been executed yet, so even if the measured value is above or
below 21 percent, no problem occurs.

(5) After the measured value has stabilized, press the [Enter] key to display the “span-

calibration complete” display shown in Figure 7.21. At that point, the measured
value is corrected to equal the span-gas concentration setting. Close the span-gas
flow valve. The valve lock-nut should be tightened completely so that the span gas
does not leak.

Enter

Span calibration

Close the span gas valve.

Zero calibration

End

Manual calibration

F7.21E.EPS

Enter

Check the zero gas

concentration value

Next

Cancel calibration

Change value: 1.00%

Manual calibration

F7.22E.EPS

Figure 7.21 Span Calibration Figure 7.22 Zero-gas Concentration

Complete Zero calibration Check Display
start Display (in Manual Calibration)

(6) Select Zero calibration as in Figure 7.21 to display the zero-gas concentration check

display (Manual calibration) . Check that the zero-gas oxygen concentration value
and the calibration gas oxygen concentration value agree. Then select Next as in
Figure 7.22.