3 fm explosionproof type – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 54

Advertising
background image

IM 11M13A01-02E

3-3

3. Installation

• Take care not to generate mechanical sparking when accessing to the

detector and peripheral devices in hazardous location

Note 4: Maintenance and Repair

• The detector modification or parts replacement by other than authorized

representative of Yokogawa Electric Corporation is prohibited and will void
CENELEC ATEX (KEMA) Flameproof Certification.

Note 5: Cable Entry

• The type of cable entry is marked near the cable entry port according to

following codes.

Type of Cable entry : Code
M20 x 1.5

:

M

1/2NPT

:

A

• The blanking elements shall be of certified flameproof type, suitable for the

conditions of use and correctly installed.
Certified cable glands that meet or exceed the requirements for EEx dIIB +
H

2

IP66, provide at least 6 threads engaged when installed, and resist heat so

that they can be used in the operating environment, should be used.

3.1.3

FM Explosionproof Type

ZR22S-B Detector for use in hazardous locations:

Note 1: Explosionproof for Class I, Division 1, Groups B, C and D

Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure Rating:

NEMA 4X

Temperature class:

T2

Note 2: Wiring

• All wiring shall comply with National Electrical Code ANSI/NEPA 70 and

Local Electrical Code.

• In hazardous location, wiring shall be in conduits as shown in the figure.

WARNING: SEAL ALL CONDUITS WITHIN 18 INCHES OF THE

ENCLOSURE.

F3.02E.EPS

Non-hazardous Location Equipment
ZR402G or AV550G

Hazardous Locations Division 1

18 inches (475mm) MAX.

Conduit

Sealing Fitting

ZR22S Detector

Figure 3.2

Note 3: Operation

• Keep the "WARNING" label to the Detector.WARNING: OPEN CIRCUIT

BEFORE REMOVING COVER. USE AT LEAST 150

Њ C HEAT

RESISTANT CABLES.

• Take care not to generate mechanical sparking when accessing to the

detector and peripheral devices in hazardous location

Note 4: Maintenance and Repair

• The detector modification or parts replacement by other than authorized

representative of Yokogawa Electric Corporation is prohibited and will void
Factory Mutual Explosionproof Approval.