IM 11M13A01-02E

1-2

1.1

< EXAxtZR > System Configuration

The system configuration should be determined by the conditions; e.g. whether calibra-
tion is to be automated, and whether flammable gas is present and requires safety
precautions. The system configuration can be classified into two basic patterns as
follows:

1.1.1

System 1

This system is for monitoring and controlling oxygen concentration in the combustion
gases of a large-size boiler or heating furnace. Clean (dry) air (21% O

2

) is used as the

reference gas and the span gas for calibration. Zero gas is fed in from a cylinder during
calibration. The gas flow is controlled by the ZA8F flow setting unit (for manual valve
operation).

~

ZA8F flow setting unit

EXA ZR402G

ZR22S Separate type Explosionproof

ZR402G Converter

Reference gas

Calibration gas

Stop valve

or

Check valve

Needle

valve

flowmeter

Instrument air

Air Set

Pressure regulator

Zero gas
cylinder

F1.1E.EPS

Span gas (*)

* Calibration gas unit same as for zero gas.

Zirconia Oxygen Analyzer, Detector

100 to
240 V AC

Non-hazardous area

Hazardous area

Figure 1.1 System configuration 1