3 installation of the converter, 1 location, 2 mounting of the converter – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 62: Warning

Advertising
background image

IM 11M13A01-02E

3-11

3. Installation

3.3

Installation of the Converter

WARNING

Converter (Model ZR402G) must not be located in hazardous area.

3.3.1

Location

The following should be taken into consideration when installing the converter:

(1) Readability of the indicated values of oxygen concentration or messages on the

converter display.

(2) Easy and safe access to the converter for operating keys on the panel.
(3) Easy and safe access to the converter for checking and maintenance work.
(4) An ambient temperature of not more than 55

Њ C and little change in temperature

(recommended within 15

Њ C in a day).

(5) The normal ambient humidity (recommended between 40 to 75% RH) and without

any corrosive gases.

(6) No vibration.
(7) Near to the detector.
(8) Not in direct rays of the sun. If the sun shines on the converter, prepare the hood (/

H) or other appropriate sunshade.

(9) Non-hazardous location.

3.3.2

Mounting of the Converter

The converter can be mounted on a pipe (nominal JIS 50A: O.D. 60.5 mm), a wall or a
panel. The converter can be mounted at an angle to the vertical, however, it is recom-
mended to mount it on a vertical plane.

Mount the converter as follows.

<Pipe Mounting>
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for

mounting the converter. (Converter weighs approximately 6 kg.)

(2) Mount the converter on the pipe. Fix it firmly on the pipe in the procedure described

in Figure 3.8.