3. Installation

3.1.7

Installation of the Detector

CAUTION

• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop

the detector, as impact will damage it.

• A gasket should be used between the flanges to prevent gas leakage. The gasket

material should be heatproof and corrosion-proof, suited to the characteristics of the
measured gas.

The following should be taken into consideration when mounting the general-use
detector:

<General-use detector>
(1) Make sure that the cell mounting screws (four) at the probe tip are not loose.
(2) Where the detector is mounted horizontally, the calibration gas inlet and the refer-

ence gas inlet should face downward.

3.1.8

Installation of the Probe Protector (ZO21R)

<Detector with a probe protector (Model ZO21R-L-

ᮀᮀᮀ- *B for dust wear

protect>

The detector is used with a probe protector to prevent the sensor from being worn by
dust particles when there is a high concentration of dust and gas flow exceeds 10m/s
(fine-carbon boiler or fluid-bed furnace).

(1) Put a gasket (provided by the user) between the flanges, and mount the probe

protector in the probe insertion hole. The probe protector should be installed so that
the notch is downstream of the measurement gas flow.

(2) Make sure that the cell assembly mounting screws (four) at the probe tip are not

loose.

(3) Where the detector is mounted horizontally, the reference gas and calibration gas

inlet should face downward.

Direction of the measurement gas flow

Mount the protector so that the notch is
on the downstream side of the gas flow.

Reference gas inlet
Calibration gas inlet

Unit: mm

F3.05E.EPS

Probe top

1050

Gasket (t1.5)

Figure 3.5 Mounting of detector with a probe protector