5. Wiring

5.1.2

Wiring

Connect the following wiring to the converter. It requires a maximum of eight wiring
connections as shown below.

(1) Detector output (connects the converter with the detector.)
(2) Detector heater power (connects the converter with the detector.)
(3) Analog output signal
(4) Power and ground
(5) Contact output
(6) Operation of the solenoid valve of automatic calibration unit
(7) Contact input

1

FG

2

AO1

(+)

3

AO1

(-)

4

AO2

(+)

5

AO2

(-)

6

CJ
(+)

7

CJ

(-)

8

TC

(+)

9

TC

(-)

10

CELL

(+)

11

CELL

(-)

12

FG

13

DI-1

14

DI-2

15

DI-C

16

AI

(+)

17

AI
(-)

18

AC-

ZERO

19

AC-

SPAN

20

AC-

COM

21

FG

22

FG

CELL

(+)

CELL

(-)

TC

(+)

TC

(-)

CJ
(+)

CJ

(-)

23

DO-1

24

DO-1

25

DO-2

26

DO-2

27

DO-3

28

DO-3

29

DO-4

30

DO-4

31

HTR

32

HTR

33

L

34

N

35

G

36

FG

HTR

HTR

100-240V AC
50/60 Hz

F5.2E.EPS

Model ZR402G Separate type Zirconia Oxygen Analyzer/
Converter

Model ZR22S Separate type Zirconia
Oxygen Analyzer, Detector

Analog

output 2

Analog

output 1

4-20 mA DC

4-20 mA DC

Digital output

Contact

input 2

Contact input 1

Contact

output 1

Contact

output 2

Contact

output 3

Contact

output 4

Solenoid valve for span gas
for automatic calibration

Solenoid valve for zero gas
for automatic calibration

Flow setting unit

*1

*1

The protective grounding for the converter should be connected to either the protective ground terminal in the equipment or
the ground terminal on the case.
Standard regarding grounding: Ground to earth, ground resistance: 100

or less.

*1

Figure 5.2 Wiring connection to the converter