2 piping for the calibration gas, 3 piping for the reference gas inlet, 4 piping for the reference gas outlet – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 73

Advertising
background image

IM 11M13A01-02E

4-4

4.1.2

Piping for the Calibration Gas

This piping is to be installed between the zero gas cylinder and the ZA8F flow setting
unit, and between the ZA8F flow setting unit and the ZR22S detector.

The cylinder should be placed in a calibration gas unit case or the like to avoid any
direct sunlight or radiant heat so that the gas cylinder temperature does not exceed 40

Њ C.

Mount a regulator valve (specified by YOKOGAWA) on the cylinder.

Mount a check valve or stop valve (specified by YOKOGAWA) on the nipple (found on
the open market) at the calibration gas inlet of the detector as illustrated in Figure 4.2.
(The check valve or the stop valve may have been mounted on the detector when
shipped.) Connect the flow setting unit and the detector to a stainless steel pipe 6 mm
(O.D.) ~ 4 mm or larger (I.D.) (or nominal size 1/4 inch).

F4.2E.EPS

Nipple

Stop valve

Tube
connection
joint

Check valve

Piping for the Calibration gas,
6 mm (O.D.) by 4 mm (I.D.)
Stainless steel pipe

Piping for the Reference gas inlet,
6 mm (O.D.) by 4 mm (I.D.)
Stainless steel pipe

Piping for the Reference gas outlet,
6 mm (O.D.) by 4 mm (I.D.)
Stainless steel pipe

Figure 4.2 Piping for the Calibration Gas Inlet

4.1.3

Piping for the Reference Gas Inlet

Reference gas piping is required between the air source (instrument air) and the flow
setting unit, and between the flow setting unit and the detector.

Insert the air set next to the flow setting unit in the piping between the air source and
the flow setting unit.

Use a 6 mm (O.D.) ~ 4 mm or larger (I.D.) (or nominal size 1/4 inch) stainless steel
pipe between the flow setting unit and the detector.

4.1.4

Piping for the Reference Gas Outlet

If ZR22S is exposed to rain or water splash, connect the pipe outlet on downward.