3 adjust lcdscreen contrast – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 183

Advertising
background image

IM 11M13A01-02E

11-12

11.2.3

Adjust LCD screen contrast

An LCD is built in the ZR402G converter. The contrast of this LCD is affected by its
ambient temperature. For this reason, the LCD is shipped, after adjusting the contrast so
as to become the most suitable in a room temperature(20-30

؇ C). However, when display

on the LCD is hard to see, adjust the LCD contrast by change the resistance of the
variable resistor; its position is shown in Fig. 11.8

Variable resistor

Figure 11.8

Advertising