Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 34

Advertising
background image

IM 11M13A01-02E

2-15

2. Specifications

Calibration:

Method;

zero/span calibration

Calibration mode; automatic, semi-automatic and manual (All are operated

interactively with an LCD touchpanel). Either zero or span can be
skipped.

Zero calibration-gas concentration setting range: 0.3 to 100 vol% O

2

(in increments

of 0.01 vol% O

2

in smallest units).

Span calibration-gas concentration setting range: 4.5 to 100 vol% O

2

(in increments

of 0.01 vol% O

2

in smallest units).

Use nitrogen-based gas mixture containing 10 vol% O

2

, and 80 to

100 vol% O

2

for standard zero-gas and standard span-gas

respectively.

Calibration period; date/time setting: maximum 255 days

Model and Codes

ZR402G

Converter
thread

-P
-G
-M
-T

G1/2
Pg13.5
M20x1.5 mm
1/2NPT

T2.4E.EPS

Separate type Zirconia Oxygen Analyzer,
Converter

Model

Suffix code

Option

code

Description

Tag plates

*1

Specify either /SCT or /PT option code.

*2 Sun shield hood is still effective even if scratched.

-J
-E
-G
-F

Display

-A

-J
-E

/H

Instruction manual

Options

/SCT
/PT

Hood (*2)

Japanese
English

Japanese
English
German
French

Always -A

Stainless steel tag plate (*1)
Printed tag plate

(*1)