2-10

2.3.2

ZO21P-H Adapter for High-Temperature Probe

The probe adapter is used to lower the sample gas to a temperature below 700

ЊC (below

300

ЊC at probe adapter surface) before it is fed to the detector.

Insertion length:

1 m, 1.5 m

Material in Contact with Gas:

SUS 316 (JIS), Zirconia, SiC or SUS 310S, SUS 304

(JIS) (flange)

Probe Material:

SiC, SUS 310S (JIS)

Installation:

Flange mounting (FF type or RF type)

Probe Mounting Angle:Vertically downward within

؎5؇

Where the probe material is SUS 310S, horizontal mounting is
available.

Construction:

Non explosion-proof. Rainproof construction

Weight:

Insertion length of 1.0 m:

approx. 6.5 kg (JIS) / approx. 8.5 kg

(ANSI)
Insertion length of 1.5 m:

approx. 7.5 kg (JIS) / approx. 9.5 kg

(ANSI)

Sample gas temperature: 0 to 1400

؇ C (when using SiC probe) 0 to 800؇ C (when using

SUS 310S probe adapter)

Sample gas pressure:

-0.5 to + 5 kPa. When using in the range of 0 to 25 vol% O

2

or

more, the sample gas pressure should be in the range of -0.5 to +0.5
kPa. (Where the sample gas pressure for the high-temperature probe
is negative, an auxiliary ejector is necessary.)

-100
-150

*A

-J
-N
-M
-L
-A
-R
-Q
-T
-S
-E

-H

-A
-B

T2.3E.EPS

Model

Suffix code

Option

code

ZO21P

Material

Insertion
Length

Flange

High Temperature Probe Adapter

SiC
SUS 310S

1.0 m
1.5 m

JIS 5K 50 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
ANSI Class 150 4 RF SUS304
ANSI Class 150 2 1/2 RF SUS304
ANSI Class 150 3 RF SUS304
JPI Class 150 3 RF SUS304
JPI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304

Style A

Description

Style code

Note: For this high-temperature use probe adapter, be sure to

specify the ZR22S probe of its insertion length 0.15 meters.