IM 11M13A01-02E

12-4

CAUTION

• Measure the thermocouple resistance value after the difference between the detector

tip temperature and the ambient temperature decreases to 50

؇C or less. If the thermo-

couple voltage is large, accurate measurement cannot be achieved.

12.1.2.3

Error-3: A/D Converter Failure/Error-4: Writing-to-memory Failure

• A/D Converter Failure

It is suspected that a failure has occurred in the A/D converter mounted in the electrical

circuits inside the converter.

• Writing-to memory Failure

It is suspected that a failure has occurred in an operation writing to the memory

(EEPROM) mounted in the electrical circuits inside the converter.

<Locating the failure, and countermeasures>

Turn off the power to the converter once and then restart the converter. If the

converter operates normally after restarting, an error might have occurred due to a
temporary drop in the voltage (falling below 85 V, the least amount of voltage required
to operate the converter) or a malfunction of the electrical circuits affected by noise.
Check whether or not there is a failure in the power supply system or whether the
converter and detector are securely grounded.

If the error occurs again after restarting, a failure in the electrical circuits is suspected.

Consult the service personnel at Yokogawa Electric Corporation.