IM 11M13A01-02E

8-16

8.6

Other Settings

8.6.1

Setting the Date-and-Time

The following describe how to set the date-and-time. Automatic calibration or blowback
works following this setting.

Proceed as follows:
(1) Press the Setup key in the Basic panel display to display the Execution/Setup

display.

(2) Select Setup in the Execution/Setup display to display the “Commissioning” (Setup)

display.

(3) Select Others in the “Commissioning” (Setup) display. The Others display then

appears, as shown in Figure 8.13.

(4) Select Clock. The Clock display then appears, as shown in Figure 8.14.
(5) Select Set date to display the numeric-data entry display. To set the date June 21,

2000, enter 2106000 and press the [Enter] key. The display then returns to the one
shown in Figure 8.14.

(5) Select Set time. Enter the time on a 24-hour basis. To enter 2:30 p.m., type in 1430

in the numeric-data entry display. Press the [Enter] key, and the time starts at 00
second.

Clock

Averaging

Fuel setup

Unit setup

Passwords

Defaults

Enter

Others

F8.13E.EPS

Enter

Set date: 1 3 / 0 7 / 0 0

Set time: 1 4 : 3 9

1 3 J ul. 2 0 0 0

1 4 : 3 9 : 5 5

Clock

F8.14E.EPS

Figure 8.13 Other Settings Figure 8.14 Clock Display

8.6.2

Setting Periods over which Average Values Are Calculated and Periods over
which Maximum and Minimum Values Are Monitored

The equipment enables the display of oxygen concentration average values and maxi-
mum and minimum values under measurement (see Section 10.1.1, later in this manual).
The following section describes how to set the periods over which oxygen concentration
average values are calculated and maximum and minimum values are monitored.

8.6.2.1Procedure

(1) Press the Setup key in the Basic panel display to display the Execution/Setup

display.

(2) Select Setup in the Execution/Setup display to display the Commissioning” (Setup)

display.

(3) Select Others in that display and then select Averaging in the Others display. The

averaging display shown in Figure 8.15 then appears.

(4) Choose “Set period over which average is calculated” and enter the desired numeric

value from the numeric-data entry display. To enter three hours, type in 003. The
period over which average values can be calculated ranges from 1 to 255 hours.

(5) Choose “Set period over which maximum and minimum is stored” and enter the

desired numeric value from the numeric-data entry display. To enter 48 hours, type
in 048. The allowable input ranges from 1 to 255 hours.