2 blowback, 1 mode – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 159

Advertising
background image

IM 11M13A01-02E

10-8

Enter

Display item

Trend graph

Auto return time: 0 min

Language:

Display setup

F10.7E.EPS

English

Deutsch
Francias

Figure 10.7 Display Setup Display

10.2 Blowback

Blowback Setup

The following sections describe the blowback setup procedures required for carrying out
blowback.

10.2.1

Mode

There are three blowback modes available:

(1) No function – blowback disabled.
(2) Semi-automatic – in this mode, touchpanel operations or contact input signals will

start and perform blowback operations according to a preset time and output stabili-
zation time.

(3) Automatic – performs blowback operations automatically according to a preset

interval.

These three modes each have the specific limitations described below:

• If “No Function” is selected

No blowback operations are executed.

• If “Semi-auto” is selected, semi-automatic blowback can be executed (but no calibra-

tion is made even when the automatic calibration startup time is reached.)

• If “Auto” is selected, automatic blowback can be executed in either the “Auto” or

“Semi-auto” mode.

To select the desired mode, follow these steps:

(1) Select the Setup key from the basic panel display to display the Execution/Setup

display. Then select Maintenance from the Execution/Setup display.

(2) Select the Blowback setup from the Maintenance display and select Mode. The mode

selection display as shown in Figure 10.8 then appears.

Enter

No function

Semi_Auto
Auto

Mode:

Blow back setup

F10.8E.EPS

Figure 10.8 Blowback Setup Display

Advertising