2 setting procedure, 3 default values – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 132

Advertising
background image

IM 11M13A01-02E

8-15

8. Detailed Data Setting

8.5.2

Setting Procedure

The following are set so that semi-automatic calibration starts when input contact open
is applied to “Input1.”

Proceed as follows:
(1) Press the Setup key in the Basic panel display to display the Execution/Setup display.
(2) Select Setup in the Execution/Setup display to display the “Commissioning” (Setup)

display.

(3) Select “Contact setup” in the “Commissioning” (Setup) display.
(4) Select “Input contacts” in the Contact setup display. The Input contacts display then

appears, as shown in Figure 8.11.

(5) Select Input1 in that display. The Input contacts display then appears, as shown in

Figure 8.12.

(6) Select Calibration start.
(7) Select Input1 “closed.” An Open or Closed selection display then appears.
(8) Choose Open.

Input1: No function

Input1 contact is active at:

Closed

Input2: No function

Input2 contact is active at:

Closed

Enter

Input contacts

F8.11E.EPS

Input1: No function

Input1 contact is active at:

Closed

Input2: No function

Input2 contact is active at:

Closed

Enter

Input contacts

No function

Cal. gas press. low
Range change
Calibration start
Process up set
Blow back start

F8.12.EPS

Figure 8.11 Input Contacts Display Figure 8.12 Input Contacts Display

8.5.3

Default Values

When the analyzer is delivered, or if data are initialized, contact input is disable.