Caution – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 157

Advertising
background image

IM 11M13A01-02E

10-6

0.0%O

2

25.0%O

2

10min./div

12.3%O

2

60 data items

Upper limit set

Lower limit set

Time per graduation,
calculated by the set
sampling period

Currently measured value

Sampling period

F10.5E.EPS

Figure 10.5 Plotting Graph for Sampling Period

10.1.2.3 Setting Upper and Lower Limit Values on Graph

Set upper- and lower-limit values on the graph in the following procedure:

Press Upper limit in the Trend graph display. The numeric-data entry key appears. Enter
the upper-limit value. Also enter the lower-limit value in the same way. The allowable
settings for both upper-limit and lower-limit values range from 0 to 100 percent oxygen.

10.1.2.4 Default Setting

When the analyzer is delivered, or if data are initialized, the set data are by default, as
shown in Table 10.3.

Table 10.3 Default Values for Graph Setting

Item Default

Value

Parameter Oxygen

concentration

Sampling period

30 seconds

Upper limit

25%O

2

Lower limit

0%O

2

T10.3E.EPS

CAUTION

If a rapid change in the measured value occurs during sampling, no sampled data are
plotted on the graph. Use the graph indication tentatively. Check the output current for
accurate data.