3 connection to the converter – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 86

Advertising
background image

IM 11M13A01-02E

5-7

5. Wiring

5.2.3

Connection to the Converter

To connect the wiring to the converter, proceed as follows:

(1) M4 screws are used for the terminals of the converter. Each wire in the cable should

be terminated in the corresponding size crimp-on terminal.

(2) When a rubber insulated glass braided wire is used for wiring to the detector, use a

terminal box. For wiring between the terminal box and the converter, basically use
cable, e.g. PVC sheathed PVC insulated cable, rather than individual wires.

Note

The above is to prevent moisture or corrosive gas from entering the converter and to
ground the detector without fail.