IM 11M13A01-02E

5-5

5. Wiring

5.2

Wiring for Detector Output

This wiring enables the converter to receive cell output from the detector, output from a
thermocouple and a reference junction compensation signal. Install wires that allow for
10

⍀ of loop resistance or less. Keep detector wiring away from power wiring.

Separate signal and power wiring.

CAUTION

Be sure to read Section 3.1.2 to 3.1.5 where the important information on writing is
provided.

CELL(+)

CELL(+)

CELL(-)

CELL(-)

TC(+)

TC(+)

TC(-)

TC(-)

CJ(+)

CJ(+)

CJ(-)

CJ(-)

FG

Shielded cables

Shielded cables

Detector

Converter

Non-hazardous area

CELL(+)

CELL(+)

CELL(-)

CELL(-)

TC(+)

TC(+)

TC(-)

TC(-)

CJ(+)

CJ(+)

CJ(-)

CJ(-)

FG

Detector

Converter

(2) Surface temperature of the detector's terminal box: exceeding 80

؇

C

Heat-resistant wiring

Terminal box

(1) Surface temperature of the detector's terminal box: 80

؇

C or less

F5.4E.EPS

Non-hazardous area

Hazardous area

Hazardous area

Figure 5.4 Wiring for detector output

Note

If shielded cables cannot be used between the detector and the terminal box, for
example, when heat-resistant wiring is used, locate the detector and the terminal box as
close together as possible.