Manual calibration – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 168

Advertising
background image

IM 11M13A01-02E

10-17

10. Other Functions

<Flow of span gas (air)>
The standard gas unit is used only when manual calibration is employed. Therefore, the
timing for flowing span gas (air) is included in the manual calibration flowchart de-
scribed in Section 10.5.2. For operation of the converter, see Section 7.12, earlier in this
manual.

(1) When the message “Open span gas valve” is displayed on the converter display

during calibration, plug the power cord into the power supply socket to start the
pump of the standard gas unit.

Enter

Open span gas valve.

Set flow span gas to

600ml/min.

Valve opened

Cancel calibration

Manual calibration

F10.15E.EPS

Figure 10.15 Manual Calibration Display

(2) Next, adjust the flow rate to 600

±

60 ml/min using the span gas valve “AIR” (the

flow check ball stops floating on the green line when the valve is slowly opened). To
rotate the valve shaft, loosen the lock nut and turn it using a flat-blade screwdriver.
Turning the valve shaft counterclockwise increases the flow rate.

(3) After adjusting the flow rate, tighten the valve lock nut.
(4) Select Valve opened (to start calibration) from the Manual calibration display shown

in Figure 10.15. Check the Trend graph display to see that the measured value is
stabilized. Then press the [Enter] key. The Manual calibration display shown in
Figure 10.16 appears. Disconnect the power cord to stop the pump.

Enter

Span calibration

Close the span gas valve.

Zero calibration

End

Manual calibration

F10.16E.EPS

Figure 10.16 Manual Calibration Display

<Flow of zero gas>
Cause a zero gas to flow according to the Manual calibration display shown in Figure
10.17.

Enter

Open zero gas valve.

Set flow span gas to

600ml/min.

Valve opened

Cancel calibration

Manual calibration

F10.17E.EPS

Figure 10.17 Manual Calibration Display