2 cleaning – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 182

Advertising
background image

IM 11M13A01-02E

11-11

11.

Inspection and Maintenance

To replace the fuse, follow these steps:

(1) Turn off the power to the converter for safe replacement.
(2) Remove the fuse from its holder. With the appropriate flat-blade screwdriver that just

fits the holder cap slot (Figure 11.7), turn the fuse holder cap 90

Њ counterclockwise.

By doing so, you can remove the fuse together with the cap.

Cap

Flat-blade
screwdriver

Fuse

Socket

F11.7E.EPS

Figure 11.7 Removing the Fuse

(3) Check the rating of the fuse and that it satisfies the following:
Maximum rated voltage: 250 V
Maximum rated current: 3.15 amperes
Type: Time-lag fuse
Standards: UL-, CSA- or VDE-approved
Part number: A1113EF

Place a new, properly rated fuse in the holder together with the cap, and push and turn

the cap clockwise 90+‹with the screwdriver to complete installation of the fuse.

11.2.2

Cleaning

Use a soft dry cloth to clean any part of the converter during inspection and mainte-
nance.