4 csa explosionproof type – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 55

Advertising
background image

IM 11M13A01-02E

3-4

3.1.4

CSA Explosionproof Type

ZR22S-C Detector for use in hazardous locations:

Note 1: Explosionproof for Class I, Division 1, Groups B, C and D

Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure Rating:

TYPE 4X

Temperature class:

T2

Note 2: Wiring

• All wiring shall comply with Canadian Electrical Code Part 1 and Local

Electrical Code.

• In hazardous location, wiring shall be in conduits as shown in the figure.

WARNING: SEAL ALL CONDUITS WITHIN 50 cm OF THE

ENCLOSURE.
UN SELLE DOIT ÊTRE INSTALLÊ ÁMOINS DE 50 cm
DU BÎTIER.

F3.03E.EPS

Non-hazardous Location Equipment
ZR402G or AV550G

Hazardous Locations Division 1

50 cm MAX.

Conduit

Sealing Fitting

ZR22S Detector

Figure 3.3

Note 3: Operation

• Keep the "WARNING" label to the Detector.

WARNING: OPEN CIRCUIT BEFORE REMOVING COVER. REFER

TO IM 11M13A01-02E. USE AT LEAST 150

؇C HEAT

RESISTANT CABLES.

OUVRIR LE CIRCUIT AVANT D'ENLEVER LE COUVERCLE.

UTILISEZ DES CÁBLES RÉSISTANTS Á 150

؇C

MINIMUM. VEUILLEZ VOUS RÉFÉRER AU IM
11M13A01-02E.

• Take care not to generate mechanical sparking when accessing to the

detector and peripheral devices in hazardous location

Note 4: Maintenance and Repair

• The detector modification or parts replacement by other than authorized

representative of Yokogawa Electric Corporation is prohibited and will void
Canadian Standards Explosionproof Certification.