Regarding the AV550G installation procedure, refer to IM 11M12D01-01E.

3.1

Installation of the Detector

3.1.1

Location

The following should be taken into consideration when installing the detector:

(1) Easy and safe access to the detector for checking and maintenance work.
(2) Ambient temperature of not more than 60

Њ C, and the terminal box should not

affected by radiant heat.

(3) A clean environment without any corrosive gases.
(4) No vibration.
(5) The measurement gas satisfies the specifications described in Chapter 2.
(6) No measured gas-pressure fluctuations.

3.1.2

CENELEC ATEX (KEMA) Flameproof Type

ZR22S-A Detector for use in hazardous locations:

Note 1: No. KEMA 04ATEX 2156

Type of Protection and Marking Code: EEx d IIB+H

2

Group:

II

Category:

2GD

Temperature class:

T2, T300

؇C

Enclosure:

IP66

Note 2: Wiring

• All wiring shall comply with local installation requirement.
• The cable entry devices shall be of a certified flameproof type suitable for

the condition of use

Note 3: Operation

• Keep the "WARNING" label to the Detector.

WARNING: DO NOT OPEN WHEN ENERGIZED. INSTALL IN

ACCORDANCE WITH THE INSTRUCTION MANUAL.
USE AT LEAST 150

؇C HEAT RESISTANT CABLES.