1 zr22s (0.15m) high-temperature-type detector – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 28

Advertising
background image

IM 11M13A01-02E

2-9

2. Specifications

2.3

High-Temperature Separate-type Detector and Related
Equipment

2.3.1

ZR22S (0.15m) High-Temperature Separate-type Detector

Standard Specifications
Construction:

Water-resistant

Probe length:

0.15 m

Terminal box:

Aluminium alloy

Probe material:

Probe material in contact with gas: SUS 316, SUS 316L (JIS)
(Probe), SUS 304 (JIS) (Flange), Zirconia (Sensor), Hastelloy B,
(Inconel 600, 601)

Weight:

Approx. 6 kg

Installation:

Flange mounting (The use of high-temperature detector probe
adapter ZO21P-H is necessary.)

Flange standard:

JIS 5 K 32 FF equivalent (thickness varies)

Mounting angle:

Any angle between horizontal and vertical downward (high-
temperature probe is fitted with an adapter)

Reference gas and calibration gas piping connection:

Rc 1/4 or 1/4 NPT

Ambient temperature: -20 to 60

؇ C (-20 to 150؇ C on the case surface)

Sample gas temperature: 0 to 700

؇ C (temperature at the measuring point of the

sampling gas. 0 to 750

؇ C or 0 to 1400؇ C when the probe adapter

for high temperature is used.
Temperature of the probe adapter shall not exceed 300

؇ C to protect

the gasket and avoid the bolts seizing together.

Sample gas pressure: -0.5 to +5 kPa: when used at the range of more than 0 to 25 vol%

O

2

, -0.5 to +0.5 kPa. (An auxiliary ejector is required for negative

pressure application.)

Refer to Section 2.2.1. for the explosionproof and other specifications.

Model and Code
Refer to “Model and Code” in page 2-5.

External Dimensions
Refer to the Figure in page 2-6.