5 input contact settings, 1 input contact functions – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 131

Advertising
background image

IM 11M13A01-02E

8-14

8.5

Input Contact Settings

8.5.1

Input Contact Functions

The converter input contacts execute set functions by accepting a remote dry-contact
(“voltage-free contact”) signal. Table 8.7 shows the functions executed by a remote
contact signal.

Table 8.7 Input Contact Functions

Item

Function

Calibration-gas pressure

While the contact signal is on, neither semi-automatic nor

drop

automatic calibration is possible.

Measuring range change

While contact input is On, range of Analog Output 1 is switched to

0-25% O

2

and "Range" is displayed on the screen. See Figure 8.10.

Calibration start

If the contact signal is applied, semi-automatic calibration

starts (only if the semi-automatic or automatic mode has been

setup). Calibration is started with an applied one second or more

time interval single-output contact signal. Even if a continuous

contact signal is applied, a second calibration is not repeated. If

you want to perform calibration, turn the contact signal off and

then back on.

Process upset

If the contact signal is on, heater power will be switched off.

(A one or more second time interval single-output signal is

(Unburnt gas

available as a contact signal.) If this operation starts, the sensor

detection)

temperature decreases and an error occurs. To restore it to normal,

turn the power off and then back on, or reset the analyzer.

Blowback start

If the contact signal is on, blowback starts. (A one second or more

time interval single-output signal is available as a contact signal.)

Even if a continuous contact signal is applied, a calibration is not

repeated. If you want to make a second calibration,

turn the contact signal off and then back on. (Refer to Section 10.2,

"Blowback")

T8.7E.EPS

Tag:

21.0

%O2

7.35mA -Output1
7.35mA -Output2

Range

F8.10E.EPS

Figure 8.10 Changing Measuring Range with Input Contact

Note

1. Measurement range switching function by an external contact input is available for

analog output1 only.

2. When making a semi-automatic calibration, be sure to set the semi-automatic or

automatic mode using the Calibration setup display.

3. When carrying out “blowback,” be sure to set “blowback” in the output contact setup.
4. When the unburnt gas detection signal is sent to the contact input, the converter will

cut the power supply to the heater of the detector. As a result, the heater temperature
becomes low and Error 1 or Error 2 happens.