10 checking current loop – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 110

Advertising
background image

IM 11M13A01-02E

7-11

7. Startup

7.10 Checking Current Loop

The set current can be output as an analog output.

(1) Press the Setup key on the basic panel display to display the Execution/Setup

display. Then select Maintenance in the Execution/Setup display.

(2) Select “mA-output loop check” in the Maintenance panel display to display the “mA-

output loop check” display, enabling you to check “mA-output1” and “mA-output2.”
Select the desired output terminal for current-loop checking (see Figure 7.15.1).

(3) At the time of entering the numeric-data entry display, the output current will change

to 4 mA (default value). If the desired current is entered, the corresponding output
will be provided.

mA - output 1

mA - output 2

Enter

mA - output loop check

F7.15.1E.EPS

Figure 7.15.1 “mA-output loop check” Panel Display

Advertising