Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 163

Advertising
background image

IM 11M13A01-02E

10-12

Table 10.5 Initialization Items and Default Values

T10.5E-1.EPS

Item

Data to be initialized

Default setting

Type of equipment

Not initialized

Equipment

Detector

ZR22

selection

Measurement gas

Wet gas

Display item

1st display item

Oxygen concentration

2nd display item

Current output 1

3rd display item

Current output 2

Tag name

Deleted

Trend graph

Parameter

Oxygen concentration

Sampling interval

30 seconds

Displayed data

Upper limit (graph)

25%O

2

Lower limit (graph)

0%O

2

Automatic return time

0 min.

Language

Not initialized

Mode

Manual

Calibration procedure Span - zero

Zero-gas concentration 1.00%O

2

Span-gas concentration 21.00%O

2

Calibration data

Calibration setting

Output hold time

10 min., 00 sec.

Calibration time

10 min., 00 sec.

Interval

30 days, 00 hr.

Start date

01/01/00

Start time

00:00

Mode

No function (invalid)

Blowback

Blowback setting

(Output) hold time

10 min., 00 sec.

Blowback time

10 min., 00 sec.

Interval

30 days, 00 hr.

Start date

01/01/00

Start time

00:00

Parameter

Oxygen concentration

mA-output 1

Min. oxygen concentration 0%O

2

mA-output 2

Max. oxygen concentration 25%O

2

@ @

Output damping

0%

Output mode

Linear

Warm-up

4 mA

Set value

4 mA

Current output

Maintenance

Previous value held

data

Output hold setting

Set value

4 mA

Calibration, blowback

Previous value held

Set value

4 mA

Error

Preset value held

Set value

4 mA