IM 11M13A01-02E

3-5

3. Installation

3.1.5

IECEx Flameproof Type

ZR22S-D Detector for use in hazardous locations:

Note 1: IECEx KEM 06.0006

IECEx flameproof type

Ex d IIB+ H

2

T2

Enclosure: IP66
IECEx type of protection "Dust"
Ex tD A21 IP66 T300

؇C

Enclosure: IP66

Note 2: Wiring

• All wiring shall comply with local installation requirement.
• The cable entry devices shall be of a certified flameproof type suitable for

the condition of use.

Note 3: Operation

• Keep the "WARNING" label to the Detector.

WARNING: DO NOT OPEN WHEN ENERGIZED. INSTALL IN

ACCORDANCE WITH THE INSTRUCTION MANUAL.
USE AT LEAST 150

؇C HEAT RESISTANT CABLES.

• Take care not to generate mechanical sparking when accessing to the

detector and peripheral devices in hazardous location.

Note 4: Maintenance and Repair

• The detector modification or parts replacement by other than authorized

representative of Yokogawa Electric Corporation is prohibited and will void
IECEx Flameproof Certification.

Note 5: Cable Entry

• The type of cable entry is marked near the cable entry port according to

following codes.

Type of Cable entry : Code
M20

ϫ 1.5

:

M

1/2NPT

:

A

• In case of ANSI 1/2NPT plug, ANSI hexagonal wrench should be applied

to screw in.

• The blanking elements shall be of certified flameproof type, suitable for the

conditions of use and correctly installed.
Certified cable glands that meet or exceed the requirements for Ex d IIB +
H

2

T2, Ex tD A21 IP66 T300

؇C, provide at least 6 threads engaged when

installed, and resist heat so that they can be used in the operating
environment, should be used.