5. Wiring

5.7

Wiring for ZR40H Automatic Calibration Unit

This wiring is for operating the solenoid valve for the zero gas and the span gas in the
ZR40H Automatic Calibration Unit, in a system where the calibration gas flow rate is
automatically controlled (e.g. System configuration 2). When installing this wiring,
proceed as follows:

COMMON SPAN

ZERO

to ZR402G terminal "AC-COM"

to ZR402G terminal "AC-SPAN"

to ZR402G terminal "AC-ZERO"

Wiring inlet

2-G1/2, Pg13.5, M20 or 1/2NPT

ZERO

SPAN

COMMON

Span gas solenoid valve

Zero gas solenoid valve

Terminal screw : M4

Use a 3-core shielded cable or equivalent

F5.11E.EPS

Figure 5.11 Automatic Calibration Unit

5.7.1

Cable Specifications

Use a three-core cable for this wiring.